5400L Series LCR Meter for High Current Test System

WK5400L5: 10Hz~5MHz / WK5400L8: 10Hz~8MHz / WK5400L10: 10Hz~10MHz

WK5400L5

10Hz~5MHz

WK5400L8

10Hz~8MHz

WK5400L10

10Hz~10MHz

Overview

Meter Mode, MultiBias Mode and Calibration Mode as standard. Graph Mode is an option.

UI operation interface is very simple.

Saturation I(Sat) testing can be done using curve sweeping or using MultiBias fast test mode.

Single trigger can produce both AC parameter and DC Resistance measurements displayed at the same time.

Bias Status displays the working indicators of each 3265B bias source.

Open/Short Circuit Trims, HF Compensation and Load Compensation in Boost/Normal mode, with comprehensive correction to ensure accuracy and stability.

Field upgradeable software. Technical support available which is easy to follow, fast and convenient. Made in the UK and fast local service.

WK5400 is the test system controller. The DC Bias Current source WK3265B machines are connected to the Bias Adapter of the 5400 directly with a dedicated wiring system. This makes it is easier to set up and maintain the system.

Key Features

  • Designed for high current inductance testing systems.
  • External DC bias current source up to 250A/1MHz or 300A/500kHz using 10 or 12 x 3265B DC Bias Units.
  • Three modes of operation with independent calibrations to cover all test requirements.
  • High Test Speed.
  • Basic Accuracy: 0.05%
  • Measurement Frequency: 10Hz~10MHz
  • AC signal source: 5Vrms/200mArms. Step size < 1% of setting.
  • Rdc measurements with 10uΩ resolution and drive levels up to 5V/200mA.
  • Allows up to 4 different parameters to be measured, each at up to 3 different spot frequencies at the same time, totalling 12 measurement results.
  • Automatic level control (ALC) maintains constant voltage or constant current at Device Under Test.
  • Two settings for source resistance: 25Ω and 50Ω.
  • Fixture Calibration only needed once. Does not need to be repeated if the fixture is not changed.
  • Data logger firmware option /D allows measurement and results storage in real time.
  • Accurate and stable testing down to pico Henry inductance level.
  • Accurate testing below 100mV AC Drive Level.
  • Design allows accurate testing at any testing frequency point.
  • Best in Class correlation between Boost Mode and Normal Mode measurements.
  • Multi-point Multi Bias with ΔL, high-speed testing
  • Diagnostic Status Page indicates status of each 3265 DC Bias Unit connected in the system.
  • Remote control via GPIB, LAN and USB interfaces.
  • Energy-saving smart switching mode: When operating in Standalone Mode, the 3265B bias source units automatically shut down.
  • Made in the UK and serviced globally.

Technical Specification

Function

Parameter

5400L5

5400L8

5400L10

Measurement Frequency

Minimum Frequency

10Hz

Maximum Frequency

5MHz

8MHz

10MHz

Frequency Resolution

≦1mHz

AC Drive

AC Voltage Range

10mV~5Vrms

AC Voltage Resolution

<1% of setting

AC Current Range

200uA~200mArms

AC Current Resolution

<1% of setting

Automatic Level Control(ALC)

ON or OFF

Maintains constant voltage or constant current at the Device Under Test

Source Resistance

Normal Mode : 50Ω ; Boost Mode : 25Ω

Basic Accuracy

0.05%

DC Resistance(Rdc)

DC Voltage Range

10mV~5V

DC Voltage Resolution

1mV

DC Resistance Resolution

10uΩ

Built-in DC Bias

Maximum current

Reserved temporarily

External DC bias

Maximum voltage

Reserved temporarily

External DC Bias Current using 3265B or 3265BQ* DC Bias Units

Maximum Current and Frequency

300A at 500kHz

250A at 1MHz

50A at 3MHz*

DC Bias Current Resolution

25mA

Option

Order Code

Item

Firmware:

1J5400OPG

Analysis Function (Graphs)

1J5400OPD

Data Logging. Measurement and results storing in real time

Internal Hardware:

1J5400DCBIASKIT

One interface board and cables to control 6 x 3265B DC Bias Units

/5B1

Bin Handler: Non-isolated 5V signals

/5B2

Bin Handler: Isolated 24V signals

/5B3

Bin Handler: Universal optical coupling card

External Hardware:

1J1041

300A High current two-wire DIP fixture

1J1031

High current two-wire SMD adaptor for 1J1041 fixture

1J1054

1054 DIP fixture replacement (for 1054 supplied as standard with 5400).

Related Products